This is an application for the Red Pitaya STEMlab 125-14 board (RP): a set of instruments built in the FPGA (oscilloscope, harmonic and square lock-in amplifiers, two PIDs, ramp generator and automatic lock control) to measure a system with modulation techniques and stabilize it with a feedback loop, controlled from a web page. The board is closed-hardware and open-software.
The project is hosted in: https://marceluda.github.io/rp_lock-in_pid — there you can find documentation about the use of the App, its instruments and some suggested applications.
Current version: 2.2.16 (branch master; package archive/lock_in+pid-2.2.16-32-devbuild).
This version is based on the 0.2.x series (v0.2.4), but with the FPGA design built in Vivado 2020.1,
using the project structure of RedPitaya-FPGA
(project v0.94, model Z10, the same port done in
rp_lock-in_pid_h v2.3.x).
It is intended for the Red Pitaya Ecosystem 2.0 (e.g. RedPitaya_OS_2.04; tested on 2.04-35).
Since 2026-10-15 the branch master holds the 2.x series; the 0.2.x series (FPGA with Vivado 2015.2), for the
0.9x / 1.0 ecosystems, stays in the tags v0.2.1 .. v0.2.4.
- 2.2.5: FPGA in
lock_in+pid/fpga/(Vivado 2020.1, see lock_in+pid/fpga/README.md). Oscilloscope of the Ecosystem 2.0 (trigger lock, decimation widened to 20 bits), exact bus reads of the lock module. Only the App itself is packaged (the_DEBUGand_RELOADflavours were removed). - 2.2.6: lock control entirely in the FPGA (
lock_statusat 0x2C replaceslock_feedback, see lock_in+pid/doc/lock_control.md), Capture Lock and Fire on click (lock_in+pid/doc/lock_capture.md), 32-bit lock time trigger (up to 34.36 s), trigger fields with units, Config → load time window, scope time range up to 68.7 s,py/lock_from_registers.py, and the other web UI fixes of lock_in+pid_harmonic 2.3.10-2.3.17. - 2.2.7: the minor fixes of lock_in+pid_harmonic 2.3.18 (same FPGA): lock trigger registers written from Python
shown and kept by the page, Single mode keeps a slow poll, Config → load without read-only keys, X+ up to 68.7 s,
lock_from_registers.py --timeclamped. - 2.2.8: pipeline registers in the FPGA to close timing (WNS −11.7 → −1.8 ns), references delayed 1 clock to keep the measured phase (lock_in+pid/doc/lock_latency.md).
- 2.2.9 (from lock_in+pid_harmonic 2.3.21–2.3.23): Reset zoom sets the scope to one period of the ramp (Ext Trig = Scan floor / Scan ceil), of the harmonic modulation (harmonic mod.), of the square modulation (Square mod.) or one whole memory at the current Dec, and Y to −1 … 1 V (raw mode −8192 … 8192); it ends a lock capture first, and Raw Mode rescales Y. Mouse wheel on the scope: no key zooms both axes, SHIFT only time (cursor ↔), ALT only Y (cursor ↕). Slow scans (windows ≥ 1.5 s, e.g. a 30 s ramp): no samples of an older lap of the scope memory, Capture Lock with long histories (the lock request waits for the scope), Reset zoom at 0.98 × the period; (FPGA) a time trigger armed after the ramp was stopped waits for the next scan floor.
- 2.2.10 (from lock_in+pid_harmonic 2.3.24): scope presets below Ext Trig:
in1|in2,in1|ctrl,X|Y(Xo, Yo),in1|err,ctrl|errandusr(a pair of the user: double click to choose it; saved in the browser and in the configs). They change only OscA/OscB (not the zoom nor the trigger). Picking an Ext Trig other than None sets the Trigger panel to External + Normal. - 2.2.11: Auto waits for the trigger (Ext Trig or the level of a channel) and forces an acquisition only if none comes within max(0.25 s, 1.1 × window, 1.1 × period of the Ext Trig when known) + 50 ms: with Ext Trig = Scan floor t = 0 is the scan floor in Auto too (before, Auto triggered at once, at any phase of the scan). Also: the option Don't turn on PIDs and Ramp on config loading (Config dialog) is now unchecked by default: Config → load / Load file (and the restore after a Restart) bring back the ramp (Scan enable) and the PIDs as saved, so a config saved while scanning loads scanning. Checking it keeps the previous protection.
- 2.2.12 (from lock_in+pid_harmonic 2.3.27): Config dialog: Load last config on init (checked by default, saved in the browser): the App opens with the last settings used in this browser (also in a new tab); unchecked, it opens with its defaults. Buttons Load last config and Reset to defaults restart the App with them. The Lock Control panel is now right below Outputs.
- 2.2.13 (from lock_in+pid_harmonic 2.3.28): Lock Control: three small buttons time / level / time+level between Trigger Lock and Lock Now: one click on a button and one on the graph set the trigger type and value and fire Trigger Lock. Fixed: a click with Choose from graph + Fire on click during a POST in flight lost the new values or the re-lock.
- 2.2.14 (from lock_in+pid_harmonic 2.3.29): Raw Mode: the Level of the scope trigger is in ADC counts (field, slider, line; −8192 … 8191) and the controller writes those counts to the scope (it took them as volts: Normal / Single never triggered for a level other than 0); AUTO in Raw Mode with its Y range in counts (it was clipped to ±1000).
- 2.2.15 (from lock_in+pid_harmonic 2.3.30): AUTO puts the trigger level at the middle of the signal also with a calibration offset of the input (in volts it was one offset off: above in its search, below in the level it left) or a calibrated full scale other than 1 V (level and Y range were scaled by 1 / full scale); in Raw Mode it gives a whole count and half counts are rounded as the page shows them (−2.5 was shown as −2 and written as −3).
- 2.2.16 (from lock_in+pid_harmonic 2.3.31 / lock_in+pid_plus 2.2.0): Ramp Controller (web page only): tabs
period / freq (Frequency in Hz with the nearest reachable value, Amp. p-p / Offset), Ramp period in s / ms
(whole period), Lock osc. screen (scope locked on the ramp: its rising half, the time zoom sets the ramp limits),
Reset zoom alternating whole period / rising half (Scan floor; one whole period with Ext Trig = harmonic mod. or
Square mod., no alternation), no stale Force X of
GET /data; the three ramp checkboxes in two columns, no "-0 mV" labels. This App has the triangle ramp only (no Sawtooth): Lock osc. screen always shows T/2. Also (from lock_in+pid_plus 2.2.1) the X window of the page survives aPOST /datathat timed out and was retried: the retry goes withoutxmin/xmax, an older answer does not undo a newer window, the page checks that the server took its window and sends it again, and aGET /datadoes not change the time units. Same FPGA and C sources as v2.2.15; tested on RP2 (Force X with a retried POST 21/21). See doc/Usage.md and doc/Structure.md. - See the CHANGELOG.
- Two lock-in amplifiers on the same input:
- slow harmonic lock-in (3 Hz – 49.6 kHz): in-phase and quadrature demodulation (X, Y), demodulation at an arbitrary phase (F1) and at double and triple frequency (F2, F3);
- fast square lock-in (square references up to 31.25 MHz): sqX, sqY, sqF.
- Internal harmonic and square modulation generators (references available on the outputs and on a digital pin).
- Two PID filters, for loop back stabilization.
- Ramp/Scan controller for response analysis, set by period or by frequency / amplitude / offset, with a Lock osc. screen mode that turns a time zoom of the scope into new scan limits.
- Lock control for automatic scan-stop / PID start (time, level or immediate trigger, run by the FPGA).
- Two-channel oscilloscope on any of 30 internal signals, triggered by internal events (scan floor, lock, ...), time ranges from 131 µs to 68.7 s.
- Remote control with Python scripts on the board (
lock_in+pid/py/); this App has no HTTP API.
| Document | Contents |
|---|---|
| doc/App.md | Instruments and technical data |
| doc/Usage.md | Web interface and procedures (scan, lock-in, lock), with screenshots |
| doc/Structure.md | Web page ↔ server ↔ FPGA, block diagrams |
| doc/FPGA.md | Register map |
| doc/Remote.md | Control from Python (scripts on the board, SSH client) |
| doc/Devel.md | Building, modifying and testing |
| CHANGELOG.md | Changes by version |
If you have a RP board you can install the Lock-in+PID application by copying the lock_in+pid folder (that
comes with the tar/zip package) to the /opt/redpitaya/www/apps folder (inside the RP).
The packages are built in archive/ (lock_in+pid-<version>-<build>-devbuild.{zip,tar.gz}). UnZip / UnTar the
package and execute from a terminal:
./upload_app.sh rp-XXXXXX.localReplace rp-XXXXXX.local by your RP local name or IP address. Also, you can use your own SSH client and upload the
lock_in+pid folder to the RedPitaya folder /opt/redpitaya/www/apps (file system in read-write mode, fpga.sh
executable). Then open http://rp-XXXXXX.local/lock_in+pid/.
For more information about the installing procedure, refer to: https://marceluda.github.io/rp_lock-in_pid/TheApp/install/ and doc/Devel.md.
The App was based on code from project: https://github.com/RedPitaya/RedPitaya/tree/release-v0.95/apps-free/scope
To build or modify the App you need a linux development environment with:
- the Linaro toolchain
gcc-linaro-4.9-2015.02-3or2014.11(arm-linux-gnueabihf), in/opt/linaro/, for the C controller; - Vivado 2020.1 (for the FPGA design, since version 2.2.5), in
/tools/Xilinx/Vivado/2020.1(or setVIVADO_SETTINGS=.../settings64.sh);lock_in+pid/fpga/Makefileloads it by itself in a clean environment. Vivado 2015.2 is not needed any more (settings.shloads it only if installed).
source settings.sh
LANG=C make # controller, FPGA (if changed) and packages in archive/Parts (from the root folder):
make -C lock_in+pid app # C controller (cross compiled)
make -C lock_in+pid fpga # FPGA with Vivado 2020.1 (red_pitaya.bit, fpga.bit.bin)
make clean # clean all (also: clean_app, clean_fpga in lock_in+pid)
make -C lock_in+pid upload RPIP=rp-XXXXXX.local # also upload_html, upload_bit, upload_pyRequirements, all the targets, tests and details: doc/Devel.md.
- An improved version with another PID added was developed by stefanputz in https://github.com/stefanputz/rp_lock-in_pid
- A version without square lock-in and with modulation added on output signals can be found in rp_lock-in_pid_h. The App and description is available on the project web page.
GPL-3.0 (see LICENSE). If you use this software, please cite it (see CITATION.cff): M. A. Luda, doi 10.1063/1.5080345.